This paper presents a 2 V programmable Josephson voltage standard with dual microwave frequency inputs and multiple output taps. The design provides three main features: (1) output voltages with nanovolt resolution, (2) the ability to perform a microwave frequency self-check based on a null voltage measurement, and (3) additional voltage output taps providing simultaneous 10:1 (or 5:1) divided voltage reference for resistive divider calibration. With low heat dissipation this device is well suited for implementation with a compact cryocooler as a turnkey traveling system.
About this Dataset
Title | Dual-Frequency-Bias Programmable Josephson Voltage Standard Circuit |
---|---|
Description | This paper presents a 2 V programmable Josephson voltage standard with dual microwave frequency inputs and multiple output taps. The design provides three main features: (1) output voltages with nanovolt resolution, (2) the ability to perform a microwave frequency self-check based on a null voltage measurement, and (3) additional voltage output taps providing simultaneous 10:1 (or 5:1) divided voltage reference for resistive divider calibration. With low heat dissipation this device is well suited for implementation with a compact cryocooler as a turnkey traveling system. |
Modified | 2024-07-01 00:00:00 |
Publisher Name | National Institute of Standards and Technology |
Contact | mailto:[email protected] |
Keywords | Josephson arrays , Measurement techniques , Precision measurements , Standards , Superconducting integrated circuits , Voltage measurement |
{ "identifier": "ark:\/88434\/mds2-3604", "accessLevel": "public", "contactPoint": { "hasEmail": "mailto:[email protected]", "fn": "Alain Rufenacht" }, "programCode": [ "006:045" ], "landingPage": "https:\/\/data.nist.gov\/od\/id\/mds2-3604", "title": "Dual-Frequency-Bias Programmable Josephson Voltage Standard Circuit", "description": "This paper presents a 2 V programmable Josephson voltage standard with dual microwave frequency inputs and multiple output taps. The design provides three main features: (1) output voltages with nanovolt resolution, (2) the ability to perform a microwave frequency self-check based on a null voltage measurement, and (3) additional voltage output taps providing simultaneous 10:1 (or 5:1) divided voltage reference for resistive divider calibration. With low heat dissipation this device is well suited for implementation with a compact cryocooler as a turnkey traveling system.", "language": [ "en" ], "distribution": [ { "downloadURL": "https:\/\/data.nist.gov\/od\/ds\/mds2-3604\/Fig%204%202D-Quantum%20locking%20range%20density%20plot.csv", "mediaType": "text\/csv", "title": "Fig 4 2D-Quantum locking range density plot" }, { "downloadURL": "https:\/\/data.nist.gov\/od\/ds\/mds2-3604\/Fig%205%20Frequency%20verification-voltage%20measurement.csv", "mediaType": "text\/csv", "title": "Fig 5 Frequency verification-voltage measurement" }, { "downloadURL": "https:\/\/data.nist.gov\/od\/ds\/mds2-3604\/Fig%206%20Keysight34420A-1mV%20range.csv", "mediaType": "text\/csv", "title": "Fig 6 Keysight34420A-1mV range" }, { "downloadURL": "https:\/\/data.nist.gov\/od\/ds\/mds2-3604\/Fig%207%20N11-300uV%20range.csv", "mediaType": "text\/csv", "title": "Fig 7 N11-300uV range" }, { "downloadURL": "https:\/\/data.nist.gov\/od\/ds\/mds2-3604\/Fig%203%20current%20margins.csv", "description": "Figure 3 dataset", "mediaType": "text\/csv", "title": "Fig 3 current margins.csv" }, { "downloadURL": "https:\/\/data.nist.gov\/od\/ds\/mds2-3604\/3604_README.txt", "mediaType": "text\/plain", "title": "Read me file" } ], "bureauCode": [ "006:55" ], "modified": "2024-07-01 00:00:00", "publisher": { "@type": "org:Organization", "name": "National Institute of Standards and Technology" }, "theme": [ "Metrology:Electrical\/electromagnetic metrology" ], "keyword": [ "Josephson arrays", "Measurement techniques", "Precision measurements", "Standards", "Superconducting integrated circuits", "Voltage measurement" ] }