Recommended Practices for Calibrated Millimeter-Wave Modulated-Signal Measurements

In this paper, we have demonstrated the importance of choosing the correct reference plane for applications such as over-the-air (OTA) modulated-signal measurements at millimeter-wave frequencies. We have employed a modulated-signal source at 44 GHz for this demonstration. The measurements have been performed using NIST's calibrated sampling oscilloscope and are traceable to the primary standards. The EVM values and distributions are obtained after complete uncertainty analyses. The source and oscilloscope mismatch measurements have been performed on a vector network analyzer (VNA) and are also shown here after complete uncertainty analyses. Each dataset shown in the paper has been obtained after running 1000 Monte-Carlo simulations.

About this Dataset

Updated: 2024-02-22
Metadata Last Updated: 2022-09-12 00:00:00
Date Created: N/A
Views:
Data Provided by:
digitally modulated signals
Dataset Owner: N/A

Access this data

Contact dataset owner Landing Page URL
Download URL
Table representation of structured data
Title Recommended Practices for Calibrated Millimeter-Wave Modulated-Signal Measurements
Description In this paper, we have demonstrated the importance of choosing the correct reference plane for applications such as over-the-air (OTA) modulated-signal measurements at millimeter-wave frequencies. We have employed a modulated-signal source at 44 GHz for this demonstration. The measurements have been performed using NIST's calibrated sampling oscilloscope and are traceable to the primary standards. The EVM values and distributions are obtained after complete uncertainty analyses. The source and oscilloscope mismatch measurements have been performed on a vector network analyzer (VNA) and are also shown here after complete uncertainty analyses. Each dataset shown in the paper has been obtained after running 1000 Monte-Carlo simulations.
Modified 2022-09-12 00:00:00
Publisher Name National Institute of Standards and Technology
Contact mailto:[email protected]
Keywords Digitally modulated signals , error vector magnitude , predistortion , reference planes , uncertainty analysis.
{
    "identifier": "ark:\/88434\/mds2-2796",
    "accessLevel": "public",
    "contactPoint": {
        "hasEmail": "mailto:[email protected]",
        "fn": "Paritosh Manurkar"
    },
    "programCode": [
        "006:045"
    ],
    "@type": "dcat:Dataset",
    "landingPage": "https:\/\/data.nist.gov\/od\/id\/mds2-2796",
    "description": "In this paper, we have demonstrated the importance of choosing the correct reference plane for applications such as over-the-air (OTA) modulated-signal measurements at millimeter-wave frequencies. We have employed a modulated-signal source at 44 GHz for this demonstration. The measurements have been performed using NIST's calibrated sampling oscilloscope and are traceable to the primary standards. The EVM values and distributions are obtained after complete uncertainty analyses. The source and oscilloscope mismatch measurements have been performed on a vector network analyzer (VNA) and are also shown here after complete uncertainty analyses. Each dataset shown in the paper has been obtained after running 1000 Monte-Carlo simulations.",
    "language": [
        "en"
    ],
    "title": "Recommended Practices for Calibrated Millimeter-Wave Modulated-Signal Measurements",
    "distribution": [
        {
            "downloadURL": "https:\/\/data.nist.gov\/od\/ds\/mds2-2796\/2796_README.txt",
            "description": "It provides a description of the uploaded datasets.",
            "mediaType": "text\/plain",
            "title": "README file"
        },
        {
            "downloadURL": "https:\/\/data.nist.gov\/od\/ds\/mds2-2796\/Figure2a_sourceMM_measurementRP.txt",
            "description": "Source mismatch at the measurement reference plane",
            "mediaType": "text\/plain",
            "title": "Figure2a_sourceMM_measurementRP"
        },
        {
            "downloadURL": "https:\/\/data.nist.gov\/od\/ds\/mds2-2796\/Figure2a_sourceMM_sourceRP.txt",
            "description": "Source mismatch at the source reference plane",
            "mediaType": "text\/plain",
            "title": "Figure2a_sourceMM_sourceRP"
        },
        {
            "downloadURL": "https:\/\/data.nist.gov\/od\/ds\/mds2-2796\/Figure2b_EVM_values.xlsx",
            "description": "Monte-Carlo EVM values for plotting with EVM distribution",
            "mediaType": "application\/vnd.openxmlformats-officedocument.spreadsheetml.sheet",
            "title": "Figure2b_EVM_values"
        },
        {
            "downloadURL": "https:\/\/data.nist.gov\/od\/ds\/mds2-2796\/Figure2b_histogram_measurementRP.txt",
            "description": "EVM distribution at the measurement reference plane",
            "mediaType": "text\/plain",
            "title": "Figure2b_histogram_measurementRP"
        },
        {
            "downloadURL": "https:\/\/data.nist.gov\/od\/ds\/mds2-2796\/Figure2b_histogram_sourceRP.txt",
            "description": "EVM distribution at the source reference plane",
            "mediaType": "text\/plain",
            "title": "Figure2b_histogram_sourceRP"
        },
        {
            "downloadURL": "https:\/\/data.nist.gov\/od\/ds\/mds2-2796\/Figure3_EVM_values.xlsx",
            "description": "Monte-Carlo EVM values for plotting with the EVM distribution",
            "mediaType": "application\/vnd.openxmlformats-officedocument.spreadsheetml.sheet",
            "title": "Figure3_EVM_values"
        },
        {
            "downloadURL": "https:\/\/data.nist.gov\/od\/ds\/mds2-2796\/Figure3_histogram.txt",
            "description": "EVM distribution",
            "mediaType": "text\/plain",
            "title": "Figure3_histogram"
        },
        {
            "downloadURL": "https:\/\/data.nist.gov\/od\/ds\/mds2-2796\/Figure4a_50GHz_module.txt",
            "description": "50-GHz module oscilloscope mismatch",
            "mediaType": "text\/plain",
            "title": "Figure4a_50GHz_module"
        },
        {
            "downloadURL": "https:\/\/data.nist.gov\/od\/ds\/mds2-2796\/Figure4a_67GHz_module.txt",
            "description": "67-GHz module oscilloscope mismatch",
            "mediaType": "text\/plain",
            "title": "Figure4a_67GHz_module"
        },
        {
            "downloadURL": "https:\/\/data.nist.gov\/od\/ds\/mds2-2796\/Figure4c_EVM_values.xlsx",
            "description": "Monte-Carlo means and 95% confidence limits plotted for 0dB, 10dB, 20dB channel attenuations when no fixed attenuator was inserted and when 3dB and 6dB fixed attenuators were inserted before the oscilloscope adapters",
            "mediaType": "application\/vnd.openxmlformats-officedocument.spreadsheetml.sheet",
            "title": "Figure4c_EVM_values"
        }
    ],
    "license": "https:\/\/www.nist.gov\/open\/license",
    "bureauCode": [
        "006:55"
    ],
    "modified": "2022-09-12 00:00:00",
    "publisher": {
        "@type": "org:Organization",
        "name": "National Institute of Standards and Technology"
    },
    "accrualPeriodicity": "irregular",
    "theme": [
        "Advanced Communications:Wireless (RF)"
    ],
    "issued": "2022-11-29",
    "keyword": [
        "Digitally modulated signals",
        "error vector magnitude",
        "predistortion",
        "reference planes",
        "uncertainty analysis."
    ]
}

Was this page helpful?